Application:
Measuring melting point for crystalline material, to define its purity. Mainly used for drugs, dyes, perfumes and other organic crystal substances.
Feature:
Photodetector, big LCD screen, numeric keypad. Melting curve can be displayed directly. Three samples can be measured simultaneously. Automatic calculation of initial, final and average.
Main Technical Parameter:
Item |
Content |
Melting point range |
Room temperature - 300¡æ |
Initiate temperature/ time seting |
50¡æ - 300¡æ / < 6min |
300¡æ - 50¡æ time |
< 7min |
Min. value of temperature display |
0.1¡æ |
Linearity rate for temperature lifting |
0.2 ¡æ/min, 0.5¡æ/min, 1¡æ/min, 1.5¡æ/min, 2¡æ/min, 3¡æ/min, 4¡æ /min, 5¡æ/min eight step |
Blunder of display value |
< 200 ¡æ Range: ¡À0.4 ¡æ£»200 ¡æ Ò»300¡æ Range: ¡À0.7 ¡æ |
Repetition of display value |
At rate for temperature lifting 1.0¡æ£¯min , 0.3 ¡æ |
Size of standard capillary tube |
Outer diameter ¦µ1.4mm, inner diameter ¦µ1.0mm |
Height of sample filling |
¡Ý3mm |
Power |
AC220V¡À22V,100w , 50Hz |
Dimension |
400mm¡Á280mm¡Á210mm |
Weight |
10Kg |
Exportation communication interface |
RS232 |
|